Determination of dosimetric characteristics for 125I seed source with FLUKA code 后印本

请选择邀稿期刊:
Краткое изложение: This work determined a new M-C code (FLUKA) that can be used to calculate the dosimetric characteristics of seed sources. Dosimetric parameters (dose rate constant, radial dose function, and anisotropy function) of model 6711 125I seed source were calculated with FLUKA. The results were compared to the relative data recommended by AAPM TG43U1: dose rate constant with FLUKA was in agreement with 2.041%; radial dose functions with FLUKA for distances ranging from 0.5 cm to 10 cm, the deviation was less than 5%. Therefore the FLUKA code can be used to calculate the dosimetric characteristics of seed sources.